ISOQX® LAMP pH Sensitive Dyestuff Kit

Read LAMP amplification by eye: positive reactions turn orange-yellow and negative reactions remain magenta. Glycerol-free Bst enzyme and UDG support colorimetric assay development and lyophilization.

Visual color readout
Isothermal amplification
dUTP / UDG system
Lyophilization compatible
RUO

Product overview

ISOQX® LAMP pH Sensitive Dyestuff Kit (LM701204). A pH-sensitive indicator is integrated into the amplification buffer for visible endpoint detection. The kit includes glycerol-free Bst II DNA Polymerase and UDGase and supports preparation of lyophilized reaction systems. Amplification is performed at 60–65°C with target-specific LAMP primers. Read results after cooling under good lighting, using suitable positive and negative controls.

Visible color contrast

Positive reactions are orange-yellow; negative reactions are magenta under the specified conditions.

Isothermal workflow

The reference amplification step is 30 minutes at 60–65°C; low-template samples may require longer.

Carryover-control chemistry

dUTP and UDGase support degradation of uracil-containing carryover templates before amplification.


LM701204 Overview Image

Specifications

ParameterValue
DetectionpH-sensitive colorimetric LAMP
EnzymeGlycerol-free Bst II DNA Polymerase
Carryover controldUTP / UDGase
Reaction volume25 µL
Buffer / enzyme per reaction10 µL 2.5× buffer / 0.5 µL enzyme
Amplification60–65°C for 30 min; low-template samples may need up to 60 min
Positive / negative colorOrange-yellow / magenta
Buffer storage / shipping−20°C / dry ice
Enzyme storage / shipping4°C / ice packs
Shelf life6 months for both components
Intended useResearch use only

Pack sizes & ordering

Catalog No.TestsBuffer / enzyme
LM701204ES011001 mL / 50 µL
LM701204ES021,00010 mL / 500 µL
LM701204ES0310,000100 mL / 5 mL

State the required pack size when ordering. Source: LM701204 handbook, pp. 2–4.

Performance & validation

The handbook describes visible endpoint discrimination between orange-yellow positive reactions and magenta negative reactions. The reference amplification step is 30 minutes at 60–65°C, with longer incubation possible for low template concentrations. No numerical limit of detection or diagnostic accuracy dataset is provided. Source: LM701204 handbook, pp. 2–4.

Read results by eye

A pH-dependent color change provides endpoint detection without a fluorescence reader. Controlled heating is still required.

30-minute amplification step

The reference workflow also includes UDG treatment and enzyme inactivation; 30 minutes is not the total procedure time.

Sample buffer matters

Strong acid/base or highly buffered samples can interfere with the indicator. Follow sample-pH and timing guidance.

Product documents

Product FAQs

Do I need an instrument?

A fluorescence reader is not required, but controlled heating is essential. A PCR instrument is strongly recommended in the handbook.

Is 30 minutes the total test time?

No. The workflow also includes pretreatment as needed, UDG treatment, inactivation and cooling.

Can I use a strongly buffered template?

High Tris-HCl concentrations can prevent color change. Follow the handbook guidance on sample pH and buffer capacity.

Is it supplied lyophilized?

No. It contains components suitable for developing a lyophilized reaction system.

Are both components stored frozen?

No. Store buffer at −20°C and enzyme at 4°C; both have a 6-month shelf life.

Is this product for diagnostic use?

No. For scientific research use only, not diagnostic or clinical procedures.